Spettroscopie VibrazionaliInfra-rossa e Raman
C. Daniel
MODULO: TECNICHE DI ANALISI
Sommario• Spettroscopia IR
– Principio della tecnica– Aspetti strumentali
• Funzionamento spettrometro
• Componenti• Parametri di misura
– Metodi di campionamento• Gas, liquido, solido
• Trasmissione, Riflettanza
• Spettroscopia Raman– Principio della tecnica– Alcuni aspetti strumentali
Why IR spectroscopy is so widelyused?
1. VERY POWERFUL TECHNIQUE
2. SPECTRA ARE (RELATIVELY) EASY TO INTERPRET3. OPERATION OF THE EQUIPMENT IS EASY
4. ALL TYPES OF SAMPLES CAN BE ANALYZED (GAS, LIQUID, SOLID)
5. FAST SPECTRUM ACQUISITION6. NO CALIBRATION
7. ALMOST NO MAINTENANCE
8. LOW COST OF BASIC SPECTROMETERS (15000 euros)
IR spectroscopyMeasurement of the amount of radiation absorbed
by a material within IR region
IR spectrumwavelength 0.8-1000 µm Frequency: 12800 - 10 cm-1
Divisions arise because of different optical materi als and instrumentation
Energia delle molecole
Come nel caso degli atomi, anche le molecole hanno una energia interna dovuta
ai diversi contributi:
Emolecola= Eelettronica + Erotazionale+ Evibrazionale + Etraslazionale + Enucleare
Eelettronica è associata agli elettroni
Enucleare è associata al nucleo
Etraslazionale è associata al movimento traslazionale della molecola (3/2 kT)
Erotazionale è associata alla rotazione della molecola su se stessa (intorno a uno o più assi),
Evibrazionale è associata al movimento reciproco degli atomi nella molecola
Rappresentazione schematica dei livelli energetici molecolari. A ogni livello elettronico sono associati diversi livelli vibrazionali e a ciascuno di questi, a sua volta, diversi livelli rotazionali
Eto
tale
= E
elet
tron
ica
+E
rota
zion
ale+
Evi
braz
iona
le
Stati elettronici Stati vibrazionali Stati rotazionali
L’energia delle molecole (come quella degli atomi) è quantizzata, per cui
sono “permessi” solo determinati livelli di energia
Ep = ½ k · x 2 = 2π2µνoscix2
Ep= Energia potenzialek= Costante di forzax = Allungamentoµ= Massa ridotta = m1. m2 / (m1 + m2) ννννosci = Frequenza dell’oscillatore
µπν k
osci 2
1=
νosc direttamente proporzionale alla costante di forza k ed inversamente proporzionale alle masse degli atomi
Si descrive una molecola biatomica come un sistema di due masse collegate da una molla
Frequenza di assorbimento
Essendo l’energia quantizzata, le transizioni saran no regolate da regole di selezione ( ∆∆∆∆n = ±1) e l’energia necessaria per la transizione ( E vib ) saràespressa dalla seguente relazione:
µπν k
osci 21=
+=21
nhE oscivib ν
Oscillatore armonico
In un oscillatore armonico l’assorbimento di energi a comporta il passaggio da un livello vibrazionale a quello immediatamente superiore ( ∆n= 1)In un oscillatore anarmonico ∆n può essere >1 anche se questo evento è poco probabile.
Le molecole reali sono meglio rappresentate da un modello di oscillatore anarmonico dove i livelli vibrazionali non sono ugualmente spaziati.
Oscillatore anarmonico
Armoniche superioriOvertones
Calcolo delle frequenze di assorbimentoLa legge di Hooke può essere applicata in maniera ap prossimata al calcolo delle frequenze assorbimento di molecole biatomiche (o di particolari legami) considerando come costanti di forza i seguenti valori:Legame singolo: k= 5 x 10 5 dyn/cm. Legame doppio: k= 10 x 10 5 dyn/cmLegame triplo: k= 15 x 10 5 dyn/cm.
Ad es. per un legame C=C :ν = 1682 cm -1 (calcolata); 1650 cm -1 (exp.)Ad es.per un legame C-H:ν = 3032 cm -1 (calcolata); 2900 cm -1 (exp.)
Masse degli atomi (⇑ massa, ⇓ ν)C-H C-C C-O C-Cl C-Br C-I3000 1200 1100 750 600 500 cm -1
Forza di legame (⇑k forza, ⇑ ν)C≡C C=C C−C 2150 1650 1200 cm -1
Ibridazione (sp > sp 2 > sp3)C-H (sp) C-H (sp 2) C-H (sp3)3300 3100 2900 cm -1
Diversi fattori influiscono sullafrequenza di assorbimento
• Le vibrazioni possono essere divise in due categorie:– Stiramento (stretching ): comportano il continuo cambiamento
della distanza di legame tra gli atomi coinvolti– Deformazione o piegamento (bending ): implicano la continua
variazione dell’angolo di legame tra gruppi di atomi
• Le vibrazioni di bending possono a loro volta essere di 4 tipi:– A forbice nel piano (scissoring )– Oscillazione nel piano (rocking )– Ondeggio fuori dal piano (wagging )– Torsione fuori dal piano (twisting )
Tipi di vibrazione
Tipi di vibrazioni: i modi di stretchingTipi di vibrazioni: i modi di stretching
Stretching simmetrico
Stretching asimmetrico
Tipi di vibrazioni: i modi di Tipi di vibrazioni: i modi di bendingbending
ScissoringRocking
Wagging Twisting
Example of vibrational modes-CH2-
Stretching modes -CH2-
asym. stretchingνas(CH2)
sym. stretchingνs(CH2)
scissoringδs(CH2)
rockingρ(CH2)
waggingω(CH2)
twistingτ(CH2)
.Bending modes -CH2-. . x
~2926 cm-1 ~2853 cm-1 ~1465 cm-1 1350-1150 cm-1~725 cm-1
Regola di selezione
Assorbimento della radiazione se la variazione di stato vibrazionale è associata ad una variazione del momento
dipolare (µ)della molecola
0≠
∂∂
rr
rµ
∂∂≈
rr
rµε Intensità banda IR aumenta con lavariazione del momento dipolare
Modi normali di vibrazioneQuante modi normali sono possibili in una molecola?
Modi normali:posizioni relative degli atomi cambiano ma la posizionee l’orientazione “medie” della molecola rimangono fisse
Una molecola ha un numero di gradi di libertà uguale al grado di libertà totaledi tutti i suoi atomi:
molecola con N atomi : 3N gradi di libertà
Esempio della CO 2
Perché non si osservano 3N-5/3N-6bande negli spettri ?
Bande aggiuntive-Bande overtones (armoniche): transizione simultanea di due o piùlivelli del medesimo modo normale (2ν, 3ν)-bande di combinazione: transizione simultanea di due livelli vibrazionali corrispondenti a due diversi modi normali (ν1 +ν2)
Bande assenti-nessuna variazione del momento dipolare-energia di vibrazioni diversi sono identiche (modi degeneri)-assorbimento troppo basso-ν fuori del range spettrale della misura
Part II
Aspetti Sperimentali
To separate IR light, a grating is used.
Grating
Light source
Detector
Sample
Slit
To select the specified IR light, A slit is used.
Spettrometrodispersivo
In order to measure an IR spectrum,the dispersion Spectrometer takesseveral minutes.Also the detector receives onlya few % of the energy oforiginal light source.
Fixed CCM
B.S.
Moving CCM
IR Light source
Sample
Detector
An interferogram is first made by the interferometer using IR light.
The interferogram is calculated and transformedinto a spectrum using a Fourier Transform (FT).
Spettromettro a trasformata di Fourier
In order to measure an IR spectrum,FTIR takes only a few seconds.Moreover, the detector receivesup to 50% of the energy of originallight source.(much larger than the dispersionspectrometer.)
Tipi di spettrometri
FTIR seminar
Interferometer
He-Ne gas laser
Fixed mirror
Movable mirror
Sample chamber
Lightsource
(ceramic)
Detector
(DLATGS)
Beam splitter
FT optical system diagram
Interferometria
La ricombinazione tra i due raggi genera un’interferenza che dipende
dal ritardo ottico tra i due raggi. Intensità misurata con il detector
Trasformata di FourierTipico Interferogramma di una sorgente IR
APPLICANDO LA TRASFORMATA DI FOURIER DELL’INTERFEROGRAMMA,SI PUO RICAVARE LO SPETTRO, B(ν) DELLA RADIAZIONE TRASMESSA DA
UN CAMPIONE POSTO DAVANTI AL DETECTOR
InterferometroEsistono diversi tipi di interferometro:-con ottiche piane (interferometro di Michelson)-con ottiche a corner cube
Michelson Corner Cube
Allineamento perenne
Attrito :Se c’è attrito nella movimentazione dello specchio mobile, vita media finita chedipende dell’uso della macchina
Rimozione attrito: Cuscino d’aria / Levitazione magnetica
Disallineamento di 1 µmVariazione interferogramma >10%
Il laserL’interferogramma viene generato mediante il campionamento
fatto con un laser He-Ne (λ = 632 nm, 15800 cm-1)
Campionamento delle intensità dell’interferogramma in un intervallo temporale, ∆t, sincronizzato con l’intensità della radiazione laser modulata dallo specchio mobile dell’interferometro.
La frequenza di campionamento è data da: laserlaser Vvf 2mod =V: velocità spostamento dello specchio mobile
DETECTOR VELOCE PERMETTE UNA VELOCITA DI CAMPIONAME NTO ELEVATA
POSSIBILE REGISTRARE NUMEROSI INTERFEROGRAMMA AL SE CONDO (ogni 20-30 µµµµs)
Componenti-SORGENTERadiazione si produce passando una corrente elettri caattraverso un elemento riscaldante
-SEPARATORE DI RAGGIO (BEAM SPLITTER)Materiale semi-riflettivo depositato su un materiale trasparente
SORGENTI PIU COMUNI PER IL MIR-Sorgente di Nernst (Misc. Zirconio, Ittrio, Torio) : 2500-500 cm-1
-SPIRA Ni-Cromo : 13000-500 cm-1
-GLOBAR (Carburo di Silicio) : 8000-250 cm-1
15000-3500Quarzo5000-200Ge su KBr
50000-1400Silicon su CaF2600-150Mylar
Intervallo di applicazione (cm-1)
Componenti-DETECTOR
-Termici: funzionano a Troom con largo intervallo di frequenza(SVANTAGGI : LENTI, SCARSA SENSITIVITA)
- Piroelettrici (cristalli piroelettrici sotili):-Deuterated Triglycine Sulfate (DTGS)-Lithium Tantalate (LITA)
- Fotoconductivi (Alta sensitività, veloce) -Mercuri Cadmium Telluride (MCT)-Indium antimonite (InSb)
1-412000-4000InGaAs10000-1350InSb
40-15010000-600MCT6-105000-200DTGS
Frequenza campionamento (kHz)
Intervallo di applicazione (cm-1)
Sensibilità Spettrale Detectors
DTGS
MCT
InSb
Generation of a FT-IR spectrum
Step 1: scanning air (empty beam) R( νννν)
Step 2: scanning the sample S( νννν)
ratio between the twosingle-beam spectra
% Transmittance
flat baseline
( ) ( )( )ννν
R
ST ×= 100
Risoluzione spettroscopica
80090010001100120013001400Numero d'onda (cm-1)
Ass
orba
nza
(u.a
.)
R = 2 cm-1
R = 4 cm-1
R = 10 cm-1
R = 20 cm-1
Capacità di risolvere due righe spettrali
Criterio di RayleighPer risolvere due righe spettrali separate da una distanza d (cm-1) si deve misurare un’interferogramma almeno su un camino L = 1/d
La risoluzione massima di uno spettrometro è uguale al camino ottico massimo consentito dall’interferometro
Parametri che influenzano la risoluzioneFUNZIONE DI TRONCAMENTOConsente di effettuare un troncamento dell’interferogramma
( ) ( ) ( )ννν BDB ⋅='
( ) ∫∆+
∆−
= δνπδδδν dDIB )2cos()()('
Perdità di risoluzione
Scelta funzione critica per:-bande strette (gas)-band fitting
Spettri in fase gasBoxcar assicura la migliore risoluzione
Misure quantitative con films polimericiBlackman-Harris
Rapporto segnale/rumore (SNR)Definito come l’inverso del livello del rumore sulla linea 100% prodotta
dividendo due single beam consecutivi misurati nelle stesse identiche condizioni
Da cosa dipende?
•Componenti ottiche (sorgente, beamsplitter, detecto r)•Parametri di misura (risoluzione, tempo di misura, velocità specchio,…)
NSNSttSNR gle ≈=≈ sin
Per aumentare di un ordine di grandezzail SNR rispetto ad una scansione bisogna
mediare 100 scan
ν∆≈SNRSNR aumenta con la risoluzione
Ideal to examine broad spectral regions andweak absorptions with high resolution
Advantages of FTIR spectroscopy• Speed
All the frequencies are recorded simultaneously; a complete spectrum can be measured in less than a second.
• SensitivityIn the interferometer, the radiation power transmitted on to thedetector is very high which results in high sensitivity.
• Internal calibrationFTIR spectrometers employ a HeNe laser as an internal wavelength calibration standard, no need to be calibrated by the user.
• Multicomponent capabilitySince the whole infrared spectrum is measured continuously, all infrared active components can be identified and their concentrations determined.
Principal types of sampling holding
TRANSMISSION SAMPLING(traditional method)
I0 I
SAMPLE
Advantages:-highest S/R ratio-No distorsion of the spectra-Quantitative analysis straightforward- Spectral librairies with trans. Meas.
Disadvantages:-Need for thin samples(liquid<2µm, film: 10-100µm)
REFLECTION SAMPLING
Io
It
Ir Detector
Advantages:- No need for thin samples
Disadvantages:-low S/R ratio-Distorsion of the spectra-Complex quantitative analysis
Transmission sample handling
GASES
Single pass cellL = 1, 4, or 10 cm
High Sensitivity (Multipass)Sample Cell L = 60 … 980 cm
LIQUIDS
Demountable liquid cellwith teflon spacers
L = 0.015 to 1 mm
Scelta delle celle
Il campione deve essere contenuto in celle che non devono assorbire le radiazioni IR nelle zone di interesse
NaCl : 625 cm -1
KBr : 400 cm -1
CsI : 200 cm -1
Cut-off
SOLUBILITA CELLA !!!!KBr : solubile in H2OAgBr : insolubile in H2O
Solid samples (powder)
i)Grind and mix with Nujol (hydrocarbon oil) to form mull
ii)Make semi-transparent pellet with KBr (1-3 mg sample in 250-300 mg KBr).
• Halide salts get transparent when pressured
CaCO3 in KBr , showing the mean diameters of the absorbing particles
55µ
40µ5µ 2µ
15µ23µ
(a) increase of light loss from reflection and scatteri ng by large particles
ARTEFACTS
(b)matching of sample and medium RI to prevent scattering.
e.g. PVC (nD = 1.548) dispersed in KBr (n D = 1.56), KCl and KI.
KBr
KCl
KI
(c) Fattori chimici o fisici (es. Reazione con Halide)
Spectra of benzoic acid in alkali halide discs.
Spettro IR del Nujol
2925 2855
1462 1377
721
Asymmetric, sloping bands. Badly ground.
Sample (mull) too thick
Sample does not cover beam.
Also for air bubble in liquid cell; polymer film with hole or crack
Possible problems with spectra
Reflectance sampling handling
Diffuse Reflectance (DRIFT)
Transmission-Reflection(absorption-reflection)
IR in IR out
IR in
Ir out
IR outIR in
Attenuated Total Reflection (ATR)
sample
Diffuse reflection by a powder Optical arrangement for DRIFT
Diffuse reflectance (DRIFT)
Applications-strongly absorbing samples -small, irregular samples, powders-surface analysis of hard objects using abrasive paperAdvantages-minimal sample preparation -sample not destroyed
Principle of the methodMeasurement of the energy that hasbeen partially absorbed, transmittedand partially scattered by a surfacewith no defined angle of reflection
Attenuated total reflection (ATR)
IR outIR in
sample
Principle of the methodwhen radiation strikes an interface with amedium of lower RI total reflection can occurfor an angle > θθθθc (sin θc = nsam / ncrys)
Radiation actually penetrates sample and is partially absorbed
BUT
Applications-strongly absorbing samples (films, pastes, gel, liquid)-Surface analysisAdvantages-Minimal sample preparation -Possible to analyzed multicoated samples
Very HardChemically inert
2.44500-2500 &1680-30
Diamond
Hard & brittle
Insoluble in water4.015500-675Ge
Insoluble in water and organic solvents
2.4320000-650ZnSe
SoftSoluble in bases
Slightly soluble in water
2.3817000-400KRS-5
general propertiesRI at 2000
cm-1
IR rangecm-1
crystal material
Choice of ATR crystal
Crystal material defines the RI, the IR range, and the chemicalproperties of the crystal
( )22sin2
1
criscampcris
P
nnnd
−=
θπν
Distance where the intensity of theevanescent wave decays to 1/e
dp is “equivalent” to the samplethickness for transmission sampling
(dp=0.5-2 µm)
Factors affecting ATR spectra
•Angle of incidence (θ):dp (and thus A) decreases with increasing θ(helpul for depth profile near the surface)
•RI of the crystal (ncris): dp (and thus A) decreases with increasing RI
•Wavenumber (ν): dp (and thus A) decreases with increasing ν(ATR spectrum differs from the analogous transmission spectrum)
• Efficiency of sample contact: dp (and thus A) decreases with
decreases the sample contact
30o
45o
60o
ATR-FTIR of China ClayFilled Polyester Film
Effetto del angolo di incidenza
Variazione angolo 30°- 60°diminuzione d diun fattore ~10.
Can depth profile by changing θθθθ.Smaller angle, deeper penetration
Celle ATR
Transmission-ReflectionIR in IR out
Principle of the methodBy placing the sample on a reflective surface, IR radiation travels through the sample, is reflected and passes a second time the sample.
Applications-coating on reflective surface (0.5-20 µm thickness range)Advantages-spectrum similar to transmission one -double pathlength
Coating of soft drink can
IR microspectrometry
• Introduced in 1980s• IR transmission or reflection spectra• Sample dimensions in 10 -500 µm• Instrument
– Ordinary optical microsocpe– FT-IR with small IR beam size– LN2 MCT (mercury/cadmium/telluride) PC
Advantages-Possible to analyze small samples-IR imaging
FTIR microscope
Transmission measurement
Cross-section of food packaging multilayer film cut to 10 µµµµm thickness by microtome.
(1), (3) are PE and PP H2O barriers
(2) Is ethylene vinyl alcohol copolymer, O 2barrier
(1) 60 µm(2) 8 µm
(3) 40 µm
(1)
(2)
(3)
Microscopic IR spectra of layers
Utilizzo di fibre otticheFibreFibre otticheottiche possonopossono essereessere usateusate per per misuremisure in in riflettanzariflettanza. . Con Con ll’’impiegoimpiego di di sondesonde èè possibilepossibile irraggiareirraggiare ilil campionecampione e e raccogliereraccogliere la la radiazioneradiazione diffusadiffusa; ;
entrambeentrambe le le radiazioniradiazioni possonopossono viaggiareviaggiare susu fibrefibre
VantagggiVantagggi ::--PossibilitPossibilitàà di di effettuareeffettuare analisianalisi in situin situ, , totalmentetotalmente non non distruttivedistruttive e e senzasenza toccaretoccare ilil campionecampione; ;
--Non Non cici sonosono vincolivincoli dovutidovuti allaalla forma del forma del campionecampione
FibreFibre otticheottiche ::vetrivetri a a calcogenurocalcogenuro (As(As--SeSe--Te) o Te) o calcoalogenurocalcoalogenuro. .
nessunanessuna delledelle fibrefibre attualmenteattualmente disponibilidisponibili èè in in gradogrado didi coprirecoprirell’’interointero range del range del mediomedio infrarossoinfrarosso; (; (massimomassimo 40004000--900 cm900 cm--11))
Sonde per riflettanza diffusa
La La sondasonda per per riflettanzariflettanza diffusadiffusa puòpuò essereesserecostituitacostituita dada unauna fibrafibra otticaottica per per ll’’irraggiamentoirraggiamento e e unauna per la per la raccoltaraccolta delladella radiazioneradiazione diffusadiffusa
Sonde ATR
NessunaNessuna limitazionelimitazione al range al range spettralespettrale analizzabileanalizzabilese non in se non in funzionefunzione del del cristallocristallo ATRATR
Caratterizzazione di film sottili mediantespettroscopia IRRAS
(Infra Red Reflection Absorption Spectroscopy)
Tecnica di campionamento deve essere adattata allo spessore dei campioni
spessore = 10 - 200 µm Trasmissione
spessore > 200 µm Riflessione Totale Attenuata (ATR), DRIFT
spessore < 1 µm IRRAS (Infra Red Reflection Absorption Spectroscopy)
IRRAS
Film sPSSpessore: 100 nm
Trasmissione
Principio di funzionamento della tecnica IRRAS
Film da analizzare deve essere depositato su una superficie
metallica riflettente
Ep
Es
due stati di polarizzazione sono definiti:•Es polarizzato ortogonalmente al piano di riflessione (parallelamente al piano del film)•Ep polarizzato parallelamente al piano di riflessione (perpendicolarmente al piano del film)
Il piano di riflessione (o d’incidenza) èindividuato dal raggio incidente e dalla
normale alla superficie riflettente
cambiamento di fase per la luce
riflessa dalla superficie metallica al
variare dell’angolo d’incidenza
E p
E s
E s
E p
Cambiamento di fase per i due stati di polarizzazione Es ed Ep:
a) stato di polarizzazione Es incidenza obliqua alla superficie
b) stato di polarizzazione Es incidenza normale alla superficie
c) stato di polarizzazione Ep incidenza normale alla superficie
d) stato di polarizzazione Ep incidenza obliqua alla superficie
•Stato di polarizzazione Es, i due vettori campo elettrico (incidente e riflesso) si annullano nel
punto di contatto � l’assorbimento della componente s è nulla.
•Stato di polarizzazione Ep, nel caso d’incidenza obliqua, i due vettori campo elettrico
(incidente e riflesso) non si annullano nel punto di contatto e la loro risultante è
perpendicolare alla superficie metallica
Coefficienti di assorbimento calcolati per gli stati di
polarizzazione Es ed Ep in funzione dell’angolo d’incidenza (θ)
(strato di acetone di 0.1 nm)
•per lo stato di polarizzazione Ep il massimo assorbimento si ha per angoli d’incidenza prossimi a 90°•per lo stato di polarizzazione Es assorbimento costante e molto basso
In sintesi:1)L’angolo di incidenza ottimale per effettuare una misura IRRAS deve essere vicino all’angolo radente (c.a. 88°)
2)Solo lo stato di polarizzazione Ep della radiazione incidente è assorbito da un film sottile o da molecole depositate su una superficie metallica.
3)La variazione del momento dipolare di transizione della molecola depositata o del film sottile deve avere una componente perpendicolare alla superficie per dare luogo ad un assorbimento della radiazione.
COMME EFFETUARE LA MISURA?
Metodo classico:1. Misura bianco con solo supporto2. Misura con film
Problema: L’intensità assorbita dal film è molto bass a����SNR basso
Alternativa:PM-IRRAS: Polarization Modulation-Infra Red Reflection Absorptio n Spectroscopy
•sorgente IR•filtro ottico•polarizzatore lineare•modulatore fotoelastico•portacampione•lente convergente•Detector (MCT)
Il modulatore fotoelastico permette di di modulare lo stato di polarizzazione della radiazione IR (stati di polarizzazione Es e Ep)
L’intensità riflessa dal campione è misurata con il detector per i due stati di polarizzazione P e S (misura simultanea):
Segnale α Ip+IsSegnale α Ip-Is
•Ip e Is sono attenuati nello stesso modo dalle molecole non orientate (H2O, CO2)•Soltanto Ip è attenuata dal film sottile
R= (Ip-Is)/(Ip+Is) permette di ottenere unicamente
l’assorbimento dovuto al film
IRRAS
PM-IRRAS
Confronto spettri IRRAS/PM-IRRAS
Polilisana(bio-polimero)
Near-infrared spectrometry
NIR shows some similarities to UV-visible spectrophotometry and some to mid-IR spectrometry. Indeed the spectrophotometers used in this region are often combined UV-visible-NIR ones.
The majority of the absorption bands observed are due to overtones (or combination) of fundamental bands that occur in the region 3 to 6 µm, usually hydrogen-stretching vibrations.
NIR is most widely used for quantitative organic functional-group analysis.
The NIR region has also been used for qualitative analyses and studies of hydrogen bonding, solute-solvent interactions, organometallic compounds, and inorganic compounds.
Bande overtones (armoniche): transizione simultanea di due o più livelli del medesimo modo normale (2 νννν, 3νννν)Bande di combinazione: transizione simultanea di du e livelli vibrazionali corrispondenti a due diversi modi normali ( νννν1 +νννν2)
14280cm-1
4000cm-1
Far-infrared spectrometry
The far-IR region is used to study low frequency vibrations, like those between metals and ligands (for both inorganic and organometallicchemistry).
The fundamental vibrations of many organometallic and inorganic molecules fall in this region due to the heavy atoms and weak bonds in these molecules.
Example: Metal halides have stretching and bending vibrations in the 650-100 cm-1 range.
Common applicationsof IR spectroscopy
-Identification of compounds by matching spectrum of unknowncompound with reference spectrum (fingerprinting)
-Identification of functional groups in unknown substances
-Identification of reaction components and kinetic studies of reactions
-Identification of molecular orientation in polymer films
- Detection of molecular impurities or additives present in amounts of 1% and in some cases as low as 0.01%
- Identification of polymers, plastics, and resins
- Analysis of formulations such as insecticides and copolymers
But also-Measurement of branching (HDPE, LDPE)
-Polymorphism in polymer (chain conformation, chain packing)- Polymer blends (compatibility, degree of mixing)
-…
Spettroscopia Raman
Effetto Raman (1928):
diffusione anelasticadella radiazione elettromagnetica dovuta all’interazione con i moti rotazionali e
vibrazionali della molecola
Effetto Raman
•IRaman<< IRayleigh(10-8)
•Ianti-Stokes<IStokes
luce di energia minore di quella richiesta per promuovere una molecola in uno stato elettronico eccitato viene assorbita dalla molecola si crea uno stato eccitato virtuale.
maggior parte della luce viene riemessa in tutte le direzioni alla stessa energia:Rayleigh scattering
L’energia di una piccola porzione di luce riemessadifferisce dalla radiazione incidente di intervalli di energie che corrispondono ad alcuni modi vibrazionali.
energie più alte (ννννo + hνννν) :la differenza di energia deriva da una perdita di energia vibrazionale della molecola (righe anti-Stokes)
energie minori (ννννo - hνννν)si è verificata una promozione vibrazionale (righe Stokes).
stokes antistokes
Spettro Raman del CCl 4
Figure is redrawn from D. P. Strommen and K. Nakamoto, Amer. Lab., 1981,43 (10), 72.
Observed in “typical”Raman
experiments
ν0 = 20492 cm-1
λ0 = 488.0 nm
Anti-Stokes lines(inelastic scattering)
-218
Raman shift cm -1
νννν0 = (ννννs - µµµµ0)
-200
Stokes lines(inelastic scattering)
-400400 200
218314
-314
-459
459
0
Rayleigh line(elastic scattering)
Ligne Stokes e anti-stokes sono simmetriche rispetto alla linea Rayleigh e la differenza in energia corrisponde alla variazione del livello vibrazionale iniziale
Spettro RamanIntensità di emissione luminosa v.s. SHIFT RAMAN ∆ν∆ν: differenza tra radiazione diffusa emessa e radiazione sorgente∆ν cade nel medio IR ed è indipendente della lunghezza d’onda della radiazione incidente
NB: Per lo spettro si prendono le linee Stokes (indi cate in cm -1 positivi)
POSIZIONE E INTENSITA delle bande sono legate ai gruppi funzionali e ai loro modi di vibrazione in maniera analoga alla spettroscopia IR.
FT-IR Flufenamic acid Aldrich as recd
0.05
0.10
0.15
0.20
0.25
0.30
Abs
FT-Raman Flufenamic acid Aldrich as recd
0
10
20
30
40
50
60
Int
500 1000 1500 2000 2500 3000 3500
Raman shift (cm-1)
GRANDI DIFFERENZE TRA LE VIBRAZIONI ATTIVE
ALL’IR E AL RAMAN
Vibrazioni di CO 2Vibrazioni di CO 2
Stiramento simmetrico Stiramento asimmetrico
IR AttivoRaman Attivo
Piegamenti
IR Attivi
++ ++--
Vibrazione IR inattiva
Polarizzabilità
Raman attivaNessun momento di dipolo
Vibrazione IR attivaMomento di dipolo oscillante
Livelli energetici di CO2Livelli energetici di CO2
667
1335
2350
20,000
Diff
eren
ze d
i ene
rgia
, c
m-1
Zero IRIR
Raman
Spettri IR e Raman di CO2Spettri IR e Raman di CO2
4000 2000 0Numeri d’onda (cm-1)
4000 2000 0
% T
rasm
issi
one
Spettro IR
Spettro Raman
Inte
nsità
diff
usa
Regola di mutua esclusione
Molecola con un centro di simmetria
nessun modo può essere sia IR che Raman attivo
Un modo vibrazionale può essere IR attivo, Raman attivo od entrambi
Esecuzione di una misura Raman
Per ottenere lo spettro Raman di un campione si utilizza unasorgente monocromatica con lunghezza d’onda nel vicino UV, nelvisibile o nel vicino infrarosso (NIR). Attualmente sono impiegatesorgenti LASER; alcuni esempi sono i seguenti:
La radiazione laser è focalizzata sul campione; le radiazionidiffuse dalla superficie sono raccolte, rivelate dal detector e mostrate sotto forma di spettro
• Laser UV, 244 o 325 nm
• Laser ad Argon, 488.0 nm o 514.5 nm
• Laser a He-Ne, 632.8 nm
• Laser a rubino, 694.3 nm
• Laser Nd-YAG (ittrio–alluminio–granato drogato con neodimio), 1064 nm
Utilizzo di laser diversi
SPOSTAMENTO RAMAN, per un dato legame ed un determinatomodo di vibrazione, è indipendente dalla λλλλ di eccitazione (i.e. sorgente laser)
spettri Raman espressi in unità di Raman shifts sono in teoria identicicon qualunque laser vengano prodotti e sono confrontabili anche se prodotti con strumenti Raman diversi
Tuttavia utilizzo di laser diversi su uno stesso campione puo generaresegnali Raman nelle stesse posizioni ma con intensità diverse
Con i pigmenti, si verifica sperimentalmente che il laser verde (514.5 nm) è più adatto nell’ottenere spettri di pigmenti blu e verdi, mentre illaser rosso (632.8 nm) è più adatto per pigmenti rossi e gialli.
In certi casi alcuni pigmenti non forniscono alcun spettro s e non sono irraggiati con il laser più opportuno
Esempio effetto lasersu un dye
Scelta del laser
La frequenza del laser impiegato, e quindi l’energia irraggiata sulcampione, influenzano in maniera decisiva lo spettro Raman che sipuò ottenere. È necessario tenere conto dei seguenti aspetti:
• l’intensità di emissione Raman è proporzionale alla quartapotenza della frequenza della sorgente : in altre parole, l’emissione ottenibile con un laser UV a 244 nm (40984 cm-1) in teoria è enormemente più intensa di quella ottenibile con un laser NIR a 1064 nm (9399 cm-1)
• l’energia in gioco con laser a più alta frequenza (UV, visibile) è in grado di attivare nel campione transizioni elettroniche non desiderate che possono generare fenomeni di fluorescenza e produrre spettri di difficile lettura ; questi fenomeni sono menoevidenti con laser meno energetici come il NIR
• un altro inconveniente dei laser ad alta frequenza è il danno chepossono causare ai campioni durante l’irraggiamento, causando fotodecomposizione e quindi emissione di spett riRaman anomali
lINNEE DELLO SPETTRO PIU STRETTE (PIU FACILI DA IDENTIFICARE)
SPETTRI PIU RICCHIVANTAGGI
IR RAMANEFFETTO FISICO ASSORBIMENTO
VARIAZIONE DEL MOMENTO DI DIPOLO(forte: legami ionici come O-H, N-H)
DIFFUSIONE ANELASTICAVARIAZIONE DELLA POLARIZZABILITA’(forte: legami covalenti come C=C, C-S)
PREPARAZIONE DEL CAMPIONE
SPESSORE OTTIMALE (TRASMISSIONE) o CONTATTO CON IL CAMPIONE (ATR)
NESSUN CONTATTO, NESSUNA DISTRUZIONE DEL MATERIALE, SEMPLICE PREPARAZIONE;H2O COME SOLVENTE O VETRO COME CELLA NON DISTURBANO LA MISURA
PROBLEMI FORTE ASSORBIMENTO DI VETRO, ACQUA, CO2
FLUORESCENZA
MATERIALI PRINCIPALMENTE COMPOSTI ORGANICI PRESSOCHE’ SENZA LIMITI
INTERVALLO DI FREQUENZE
4000 - 400 cm-1 (MIR) 4000 - 50 cm-1
CONFRONTO IR - RAMANle due tecniche sono più complementari
che competitive
PER LE SOSTANZE INORGANICHE, i segnali Raman relativi ailegami metallo-legante sono generalmente nel range 100-700 cm -1,
(regione spettrale di difficile utilizzo nell’IR)